Blue M Light Industrial Equipment
|
|
|
|
|
|



Model 59013
Larger image
|
|
Hydrogen Atmosphere Admission System
The Blue M 59013 Atmosphere Admission System is designed for
use with the Blue M retort furnace systems. The 59013 automatic
atmosphere control system provides dependable, operational convenience
and consistent processing with Blue M’s combustible atmosphere
retorts.
Features / Benefits
The Model 59013 is designed to provide semi-automated control of hydrogen
process gas and nitrogen purge gas. The system, which incorporates
an atmosphere burnoff assembly, can be used with a Blue M atmosphere
retort, process tube or other atmosphere tight enclosure. Customer
supplied piping is necessary between the gas supply, the 59013 atmosphere
control system, the retort and the burnoff assembly.
Nitrogen gas is used to purge the system before introducing hydrogen.
The timed purge period, both at the beginning and end of the hydrogen
process cycle, may be varied from 2- 10 minutes. Sequential pushbuttons
and indicating lights are used to activate and monitor the process
steps. Hydrogen and nitrogen flow rates are controlled by two direct
reading, adjustable, flowmeters.
An initial nitrogen purge begins immediately after the nitrogen supply
line is opened. A single lighted pushbutton will illuminate when the
nitrogen purge is complete and when depressed will start the hydrogen
gas flow. An additional lighted pushbutton will stop the hydrogen
flow and automatically switch to nitrogen purge.
The atmosphere burnoff assembly can be attached either directly or
piped to the selected atmosphere tight enclosure. The burnoff assembly
is equipped with two ignitor elements (one used as backup) to assure
ignition of the effluent hydrogen. Display lights and audible alarms
indicate low purge gas pressure, low atmosphere gas pressure, and
filament burn-out. In the event of power failure the hydrogen gas
flow will shut off and nitrogen purge started.
When ordered with a Blue M retort this system will automatically
actuate a retort plug lock during hydrogen gas flow and final purge
cycle. Plug locks must be ordered for the retort.
NOTE: When the Model 59013 is ordered with a Blue M temperature
control console, the two systems will be built into a two separate
cabinets unless specified otherwise. The temperature control console
should be ordered with Option –B Excess Temperature Protection
because of the hydrogen application.
Options
- Other Inert or Combustible
- Gas Flowmeters
|
|
|
| |
|
|
|
|
|
|